Title: Determination of full energy peak efficiency of HpGe detector from 59.5 to 1332.5 keV
Authors: Uludağ Üniversitesi/Fen-Edebiyat Fakültesi.
Gültekin, Ali Haydar
Kaynak, Gökay
Gürler, Orhan
AAH-1837-2021
7004173605
12042075600
14019444500
Keywords: Physics
Monte Carlo simulation
HpGe detector
Efficiency
Calibration
Samples
Simulation
Gamma-ray
Coincidence corrections
Monte-carlo method
Including self-absorption
Issue Date: Apr-2006
Publisher: Natl Inst Science Communication - Niscair
Citation: Gültekin, A. H. vd. (2006). ''Determination of full energy peak efficiency of HpGe detector from 59.5 to 1332.5 keV''. Indian Journal of Pure and Applied Physics, 44(4), 281-286.
Abstract: To detect very low radio activities in the environmental samples, it is sometimes necessary to know the full energy peak efficiency of the detector. This work proposes a full energy peak efficiency simulation which represents adequately the high purity germanium coaxial detector responses in the energy range 59.5-1332.5 keV and at distances of 5, 10 and 25 cm.