Başlık: Treatment of complex intracranial aneurysms using flow-diverting silk stents: An analysis of 32 consecutive patients
Yazarlar: Büyükkaya, Ramazan
Uludağ Üniversitesi/Tıp Fakültesi/Beyin ve Sinir Cerrahisi Anabilim Dalı.
Uludağ Üniversitesi/Tıp Fakültesi/Radyoloji Anabilim Dalı.
0000-0002-3425-0740
Kocaeli, Hasan
Yıldırım, Nalan
Cebeci, Hakan
Erdoğan, Cüneyt
Hakyemez, Bahattin
AAI-2318-2021
6603500567
36867883100
56033553000
8293835700
6602527239
Anahtar kelimeler: Cerebral aneurysms
Flow diverter
Silk stent
Endovascular treatment
Embolization device
Follow up
Experience
Reconstruction
Efficacy
Artery
Safety
Neurosciences & neurology
Radiology, nuclear medicine & medical imaging
Yayın Tarihi: 2014
Yayıncı: Sage Publications
Atıf: Büyükkaya, R. vd. (2014). "Treatment of complex intracranial aneurysms using flow-diverting silk stents: An analysis of 32 consecutive patients". Interventional Neuroradiology, 20(6), 729-735.
Özet: This study describes the pen-procedural and late complications and angiographic follow-up results of 32 patients with 34 complex aneurysms treated with flow diverter Silk stents in a single centre. In this retrospective study, 40 Silk stents (SS) were implanted in 34 complex intracranial aneurysms in 32 patients. In our series, 20 (58.8%) carotid-ophthalmic internal carotid artery (ICA), six (17.6%) cavernous ICA, two (5.9%) supraclinoid ICA, two (5.9%) petrosal ICA (the same patient-bilateral) and four (11.8%) posterior circulation aneurysms were treated. One of the posterior circulation lesions was a fenestrated-type aneurysm. Twenty wide-necked, saccular; eight neck remnant; four fusiform and two blister-like aneurysms were included in our series. SS were successfully implanted in all patients (100%). Mis-deployment occurred in 17.6% of patients. In two of these patients adequate stent openness was achieved via Hyperglide balloon dilatation. Coil embolization in addition to SS placement was utilized in four aneurysms. One patient (3%) experienced transient morbidity due to a thromboembolic event and there was one mortality (3%) due to remote intraparenchymal haemorrhage. Complete occlusion of 27/33 (81.8 %) and 29/33 (87.9 %) aneurysms was achieved six and 12 months after the procedure, respectively. In-stent intimal hyperplasia was detected in 6.1 % patients. Flow-diverter Silk stent implantation is an effective method of treating complex aneurysms with acceptable mortality and morbidity rates. Complete occlusion is achieved in most of the complex aneurysms.
URI: https://doi.org/10.15274/INR-2014-10070
https://journals.sagepub.com/doi/abs/10.15274/INR-2014-10070
http://hdl.handle.net/11452/27176
ISSN: 1591-0199
2385-2011
Koleksiyonlarda Görünür:Scopus
Web of Science

Bu öğenin dosyaları:
Dosya Açıklama BoyutBiçim 
Kocaeli_vd_2014.pdf94.96 kBAdobe PDFKüçük resim
Göster/Aç


Bu öğe kapsamında lisanslı Creative Commons License Creative Commons