Bu öğeden alıntı yapmak, öğeye bağlanmak için bu tanımlayıcıyı kullanınız: http://hdl.handle.net/11452/29487
Başlık: Determination of As, Cd, and Pb in tap water and bottled water samples by using optimized GFAAS system with Pd-Mg and Ni as matrix modifiers
Yazarlar: Bakırdere, Sezgin
Yaroǧlu, Tolga
Tirik, Nihan
Fidan, Ali Kemal
Maruldalı, Onur
Karaca, Abdullah
Uludağ Üniversitesi/Karacabey Meslek Yüksekokulu.
Demiröz, Mehmet
55670588100
Anahtar kelimeler: Biochemistry & molecular biology
Spectroscopy
Atomic-absorption-spectrometry
Solid-phase extraction
Neuronal apoptosis
Drinking-water
Soft drinks
Lead
Cadmium
Cancer
Preconcentration
Contamination
Arsenic
Cadmium
Nickel
Optimization
Palladium
Trace analysis
Certified reference materials
Certified values
Detection limits
Graphite furnace atomic absorption spectrometry
Lead concentration
Matrix modifiers
Optimized conditions
Reference material
Lead
Yayın Tarihi: 2013
Yayıncı: Hindawi Publishing
Atıf: Bakırdere, S. vd.(2013). "Determination of As, Cd, and Pb in tap water and bottled water samples by using optimized GFAAS system with Pd-Mg and Ni as matrix modifiers". Journal of Spectroscopy, 2013.
Özet: Arsenic, lead and cadmium were determined in tap and bottled water samples consumed in the west part of Turkey at trace levels. Graphite Furnace Atomic Absorption Spectrometry (GFAAS) was used in all detections. All of the system parameters for each element were optimized to increase sensitivity. Pd-Mg mixture was selected as the best matrix modifier for As while the highest signals were obtained for Pb and Cd in the case of Ni used as matrix modifier. Detection limits for As, Cd, and Pb were found to be 2.0, 0.036 and 0.25 ng/mL, respectively. 78 tap water and 17 different brands of bottled water samples were analyzed for their As, Cd and Pb contents under the optimized conditions. In all water samples, concentration of cadmium was found to be lower than detection limits. Lead concentration in the samples analyzed varied between N.D. and 12.66 +/- 0.68 ng/mL. The highest concentration of arsenic was determined as 11.54 +/- 2.79 ng/mL. Accuracy of the methods was verified by using a certified reference material namely Trace Element in Water, 1643e. Results found for As, Cd, and Pb in reference materials were in satisfactory agreement with the certified values.
URI: https://doi.org/10.1155/2013/824817
https://www.hindawi.com/journals/jspec/2013/824817/
http://hdl.handle.net/11452/29487
ISSN: 2314-4920
2314-4939
Koleksiyonlarda Görünür:Scopus
Web of Science

Bu öğenin dosyaları:
Dosya Açıklama BoyutBiçim 
Demiröz_vd_2013.pdf1.64 MBAdobe PDFKüçük resim
Göster/Aç


Bu öğe kapsamında lisanslı Creative Commons License Creative Commons