Title: | The dissection of reinforced endotracheal tube internal wall causing intraoperative airway obstruction under general anesthesia. Case report. |
Authors: | Mercanoğlu, Esra Topuz, Derya Uludağ Üniversitesi/Tıp Fakültesi/Anesteziyoloji ve Reanimasyon Anabilim Dalı. Kaya, Nur |
Keywords: | Anesthesiology Endotracheal tube Inner wall dissection Case report General anesthesia |
Issue Date: | 2013 |
Publisher: | Elsevier |
Citation: | Mercanoğlu, E. vd. (2013). "The dissection of reinforced endotracheal tube internal wall causing intraoperative airway obstruction under general anesthesia. Case report". Revista Brasileira de Anestesiologia, 63(4), 373-375. |
Abstract: | Endotracheal intubation is performed to establish a secure airway. However, this carries its risks and obstruction of an endotracheal tube (ETT) is a potentially life-threatening event. We report two cases with an obstruction of the resterilized, single use, spiral, reinforced endotracheal tubes by dissection of the internal wall. As a conclusion, we suggest not reusing and resterilizing single tubes in these cases to avoid a complication like dissection of the internal wall of the tube, as this has been the main cause. |