Başlık: The dissection of reinforced endotracheal tube internal wall causing intraoperative airway obstruction under general anesthesia. Case report.
Yazarlar: Mercanoğlu, Esra
Topuz, Derya
Uludağ Üniversitesi/Tıp Fakültesi/Anesteziyoloji ve Reanimasyon Anabilim Dalı.
Kaya, Nur
Anahtar kelimeler: Anesthesiology
Endotracheal tube
Inner wall dissection
Case report
General anesthesia
Yayın Tarihi: 2013
Yayıncı: Elsevier
Atıf: Mercanoğlu, E. vd. (2013). "The dissection of reinforced endotracheal tube internal wall causing intraoperative airway obstruction under general anesthesia. Case report". Revista Brasileira de Anestesiologia, 63(4), 373-375.
Özet: Endotracheal intubation is performed to establish a secure airway. However, this carries its risks and obstruction of an endotracheal tube (ETT) is a potentially life-threatening event. We report two cases with an obstruction of the resterilized, single use, spiral, reinforced endotracheal tubes by dissection of the internal wall. As a conclusion, we suggest not reusing and resterilizing single tubes in these cases to avoid a complication like dissection of the internal wall of the tube, as this has been the main cause.
URI: https://doi.org/10.1016/j.bjane.2012.07.009
https://www.sciencedirect.com/science/article/pii/S0104001413000158
http://hdl.handle.net/11452/32783