Başlık: Complications of brain tissue pressure monitoring with a fiberoptic device
Yazarlar: Uludağ Üniversitesi/Tıp Fakültesi/Nöroşirurji Anabilim Dalı.
Uludağ Üniversitesi/Tıp Fakültesi/Anesteziyoloji ve Reanimasyon Anabilim Dalı.
0000-0002-1190-6831
Bekar, Ahmet
Gören, Suna
Korfalı, Ender
Aksoy, Kaya
Boyacı, Suat
AAI-3551-2021
Anahtar kelimeler: Neurosciences & neurology
Surgery
Complications
Fiberoptic device
Intracranial pressure monitoring
Severe head-injury
Intracranial-pressure
Aggressive treatment
Clinical-evaluation
Ventriculostomy
Infections
Management
Risk
Yayın Tarihi: 1998
Yayıncı: Springer
Atıf: Bekar, A. vd. (1998). "Complications of brain tissue pressure monitoring with a fiberoptic device". Neurosurgical Review, 21(4), 254-259.
Özet: Seventy-five patients with intracranial hypertension whose Glasgow Coma Score (GCS) was 8 or below and in whom intracranial pressure (ICP) was monitored were examined for complications of this procedure. In 20 of the 75 patients we used only an intraparenchymal fiberoptic ICP monitoring transducer, while, in the remaining 55 patients, who required CSF drainage, a ventricular drainage set (VDS) was used in addition to ICP monitoring. The duration of monitoring with the ICP transducer alone was approximately 5.1 +/- 2.6 das (min. 1, max. 13) and that of ICP monitoring with VDS was 6.2 +/- 3.1 days (min. 1, max. 13). In 8 cases a total of 9 complications were experienced (12 %). These complications were infection in 3 cases (4 %), epidural hematoma in 2 cases (2.7 %): disconnection in 2 cases (2.7 %) and contusion in 2 cases (2.7 %). Although none of the 44 patients who were monitored for less than 5 days experienced infection, 3 of the 31 patients monitored for longer than 5 days did experience infection (9.7 %) (p < 0.05). None of the 20 patients who underwent ICP monitoring only experienced infection. However, 3 of the 55 patients in whom the ventricular drainage set was implanted in addition to the transducer for ICP monitoring experienced infection (p < 0.05). Owing to its minimally invasive nature, low complication rate, and accuracy in monitoring the parenchyma pressure, the Camino fiberoptic intraparenchymal monitor has become the system of choice in our clinic.
URI: https://doi.org/10.1007/BF01105781
https://link.springer.com/article/10.1007/BF01105781
http://hdl.handle.net/11452/21029
ISSN: 0344-5607
Koleksiyonlarda Görünür:Web of Science

Bu öğenin dosyaları:
Bu öğeyle ilişkili dosya bulunmamaktadır.


DSpace'deki bütün öğeler, aksi belirtilmedikçe, tüm hakları saklı tutulmak şartıyla telif hakkı ile korunmaktadır.