Please use this identifier to cite or link to this item:
http://hdl.handle.net/11452/33106
Title: | Does the magnetic field of a magnetic stirrer in an optical aggregometer affect concurrent platelet aggregation? |
Authors: | Sebik, Oğuz Çelebi, Gürbüz Uludağ Üniversitesi/Tıp Fakültesi/Biyofizik Anabilim Dalı. Sağdilek, Engin AAH-4397-2021 16835756600 |
Keywords: | Life sciences & biomedicine - other topics Biophysics Platelets Modified aggregometry Stir bar Frequency electromagnetic-fields Adenylate kinase-activity Rod outer segment In-vitro Blood-platelets Cells Metabolism Exposure Collagen Electromagnetic fields Magnetic field effects Plasma (human) Adenosine diphosphate Elf-emf Magnetic stir bars Magnetic stirring Magnetic-field Modified aggregometry Plasma samples Platelet aggregation Platelet rich plasma Stir bar Platelets |
Issue Date: | Jul-2013 |
Publisher: | Wiley |
Citation: | Sağdilek, E. vd. (2013). “Does the magnetic field of a magnetic stirrer in an optical aggregometer affect concurrent platelet aggregation?”. Bioelectromagnetics, 34(5), 349-357. |
Abstract: | Platelets are subjected to extremely low frequency electromagnetic fields during standard aggregometry measurements owing to the use of a magnetic stir bar in the instrument. This study evaluates the effects of this magnetic field exposure on platelet aggregation by comparing the results obtained in a modified aggregometer. Blood samples from healthy volunteers were anticoagulated using citrate or heparin. Platelet-rich plasma (PRP) samples were prepared. A mechanical stirring device was attached to the aggregometer instead of the magnetic stir bar system. The PRP samples were stirred using a stirring rod tip that did not produce any magnetic fields in one channel of the aggregometer; in the other channel, a stirring rod carrying a small magnet at its tip was used. As a result, a magnetic field in the extremely low frequency range and in the amplitude range of 1.9-65mT was applied to the platelets assigned to the channel where the magnetic stirring rod tip was used. Aggregation was induced using adenosine diphosphate (ADP), collagen, or epinephrine. The slopes, maximum aggregation values, and areas under the aggregation curves were compared between the magnetic and neutral stirring rod tip groups. For samples stirred with the magnetic stirring rod tip, a significant decrease was observed in 12 of the 14 parameters evaluated for aggregations induced with ADP or collagen compared to the neutral stirring rod tip, regardless of the method used for anticoagulation. This observation indicates that the magnetic stir bars used in standard aggregometry may significantly alter aggregation parameters and platelets may be possible targets of electromagnetic fields. |
URI: | https://doi.org/10.1002/bem.21785 https://onlinelibrary.wiley.com/doi/10.1002/bem.21785 http://hdl.handle.net/11452/33106 |
ISSN: | 0197-8462 |
Appears in Collections: | Scopus Web of Science |
Files in This Item:
There are no files associated with this item.
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.